15d. Transhumeral Harnessing Variations

Transhumeral harnesses are individually designed to maximize the efficiency of the control system, the suspension of the prosthesis and ideally fit comfortably on the person. Variations on harness designs may include any of the elements from the following: Hessing pad, Shoulder Saddle and Chest Strap, Triple Control Harness, Bilateral harness and Shoulder Disarticulation Harness designs.

In addition, material selection is important. Most harnesses are fabricated with polypropylene webbing, but other options like climbing webbing may be used if increased tension strength is necessary. Finally, buckles might be used to ease one-handed application/removal such as fidlock, click medical or others, e.g., WBC Industries. Be creative in your design. During your fitting appointment, consider these questions:

  • How does the location and path of the control attachment strap affect the efficiency of force and excursion?
  • How does the force on the strap and thus the cable affect the socket fit on the residual limb?
  • How do the suspension straps resist gravity and the weight of the prosthesis throughout shoulder ROM?
  • How to improve comfort?

Ultimately, the key points on harnessing must be adhered to:

  1. The harness suspends the prosthesis by distributing the load and maintains prosthetic stability through various ROM;
  2. The control system allows for the transmission of power from the individuals body motion to the prosthesis;
  3. Harness and cable system together should be operable with relatively inconspicuous body motions.
  4. The harness must be easy for the person to don, doff, launder; and
  5. It must be comfortable and durable.

Variations[1]

  1. Hessing Pap
  2. Shoulder Saddle and Chest Strap
  3. Triple Control Harness
  4. Bilateral Transhumeral Harness
Hessing Pad

The Hessing pad can replace axilla loop: Fits under scapula on contralateral axilla

Purpose: to distribute and offload weight/pressure from the contralateral axilla

Located laterally on chest wall under the scapula from about T7-T11 ribs. Notice that it does not extend very far medially.

Materials: Leather, Plastic, Spenco

Reference: Fillauer

 

Shoulder Saddle and Chest Strap

The greatest disadvantages of the chest strap harness are difficulty in applying and removing, and a tendency to rotate about the chest because of lack of a stable anchoring point.  In addition, many amputees, particularly women, cannot tolerate the chest strap.

In general, the use of the chest strap harness is confined to those patients who must lift heavy loads, or who cannot tolerate the axilla loop, even with a pad, or who have habitually worn the chest strap harness and do not wish to change.  (NYU 1979). The control attachment strap is usually connected to the lateral corner of the posterior portion of the shoulder saddle.

  • Think about reaction points!
  • Compare this design with the transradial shoulder saddle harness.
  • Note that only one lateral suspensor loop is used, connected to the socket by a D-ring, through which it is free to slide.
Strap saddle harness

 

 

 

Triple Control Harness

The NYU Triple Control Harness consists of 3 control elements:

  • Scapular abduction to operate the TD
  • Arm flexion to flex the forearm
  • Arm extension to lock and unlock the elbow unit

Consider challenges with separation of control for the patient.

 

Bilateral Transhumeral Harness

The bilateral harness is a combination of two above-elbow Figure-8 harnesses; each axilla loop is replaced by an attachment to the opposite prosthesis.  A cross-back strap and two over-the-shoulder straps may be added to improve harness stability.  The dual control system used on each prosthesis is identical with that used on a unilateral above-elbow prosthesis.

  • 2 figure-of-eight harness without axilla loops
  • Control attachment strap is continuous with the anterior support strap
  • Intersection is at midline of back
  • Lateral support straps consist of a continuous piece of nylon webbing passing over acromion and attaching to the lateral proximal aspect of socket just anterior to acromion processes. Posteriorly, stitch to anterior support straps.
  • Cross-back strap runs horizontally between the control attachment straps
  • Over-the-shoulder straps connect the control attachment strap to the lateral support strap and the anterior support strap.

The fairlead control cable system is used in the above-elbow prosthesis to transmit force for two functions: elbow flexion and terminal device operation. A second cable controls elbow locking and unlocking: the elbow-lock cable. When the elbow is extended and unlocked, flexing the shoulder transmits force to the forearm lever loop, raising the forearm to the desired level. If the user wishes to use the TD at this point, he/she locks the elbow and then operates the TD by continuation of the control motion, i.e. shoulder flexion.

 

Control motions:

  • Humeral flexion and/or scapular abduction operate elbow flexion and TD
  • Humeral extension, abduction and shoulder depression operate the elbow lock
  • Difficulty operating both prostheses simultaneously
  • Limits weight limit of either prosthesis

 